BS EN 60645 — Electroacoustics Audiometric Equipment Standard Series

Electroacoustics — Audiometric Equipment(IEC 60645 / EN 60645 Series)

🩺 BS EN 60645 is the British Standard series that adopts the European Standard EN 60645 (identical to the IEC 60645 series) for electroacoustics audiometric equipment — the “audiometer” standards. It specifies technical requirements, performance limits and test methods for pure-tone audiometry equipment, short-duration test signals and aural acoustic impedance/admittance instruments. It is the foundational standard for hearing diagnosis, occupational noise hearing screening and hearing-aid fitting. Batch BSB053 covers five documents: 60645-1, -3, -5, -6 and -7.

💡 The “hearing level” reported by an audiometer must be traceable to a standard: the 60645 series defines equipment frequency, sound-level accuracy and calibration conditions so that hearing thresholds measured by different hospitals and different audiometer models are comparable — the prerequisite for reliable hearing diagnosis.

1. Scope and General Framework 📌

Audiometric equipment quantitatively measures the hearing threshold and auditory function of the human ear. The 60645 series specifies requirements by equipment type:

  • 60645-1 (2015): equipment for pure-tone audiometry — frequency range, level range, level steps, distortion and signal characteristics; the core standard for air- and bone-conduction pure-tone audiometers.
  • 60645-3 (2007): test signals of short duration — time- and frequency-domain characteristics of pulsed tones and short tones.
  • 60645-5 (2005): instruments for the measurement of aural acoustic impedance/admittance — measurement range and accuracy of tympanometric instruments.
  • 60645-6 (2010): instruments for short-duration signals used in limited audiometry.
  • 60645-7 (2010): signal characteristics of average hearing-threshold measurement systems (for occupational/population screening).
⚠️ Documents 60645-6 and -7 in this batch are scanned versions, written from public-domain knowledge: Part 6 covers short-duration signals for limited audiometry and Part 7 covers signal characteristics of average-hearing-threshold measurement systems for group screening. The points are summarised from well-established practice; specific clauses are governed by the official BSI text.

2. Series Structure and Sub-standard Overview 🗂️

StandardSubjectEditionCore content
BS EN 60645-1:2015Pure-tone audiometry equipment2015, replaces 2001 & 60645-4:1995Frequency, level, distortion and signal requirements for air/bone audiometers
BS EN 60645-3:2007Short-duration test signals2007, replaces 1995Time/frequency characteristics of pulsed and short tones, reference figures
BS EN 60645-5:2005Aural impedance/admittance instruments2005, replaces BS EN 61027:1993Measurement range and accuracy of tympanometric instruments
BS EN 60645-6:2010Short-duration signals for limited audiometry2010Short-duration signal requirements for limited audiometry (scanned)
BS EN 60645-7:2010Average hearing-threshold systems2010Signal characteristics of screening systems (scanned)

3. Key Technical Requirements 🔧

3.1 Pure-tone audiometry equipment (60645-1)

The pure-tone audiometer is the “gold standard tool” of hearing diagnosis. The 2015 edition specifies that the equipment shall cover the stated frequency range (e.g. 125 Hz to 8 kHz, extensible higher) and hearing-level range, with the level stepped in 5 dB increments, and meet requirements for frequency accuracy, level accuracy, harmonic distortion limits and signal rise/fall time (envelope). It covers both air- and bone-conduction outputs, and specifies masking noise, automatic audiometry procedures and signal-switching characteristics.

3.2 Short-duration test signals (60645-3)

Short-duration signals (pulsed tones, short tones) are used for special hearing tests such as acoustic-reflex thresholds and recruitment tests. The 2007 edition specifies their time envelope, rise/fall time, duration, frequency content and maximum level, and updates reference figures and definitions so different instruments produce “comparable” short-tone stimuli.

3.3 Aural impedance/admittance instruments (60645-5)

Middle-ear function tests (tympanometry) apply a probe tone to the ear canal and measure the reflected acoustic impedance/admittance. The 2005 edition (replacing BS EN 61027:1993) specifies probe-tone frequency and sound pressure level, pressure range, equivalent-volume measurement range and accuracy, covering typical conditions such as 0.5 Hz to 2 kHz probe frequencies and ±400 daPa pressure sweeps, standardising tympanograms, static compliance and acoustic-reflex threshold measurements.

🚨 Periodic calibration is essential: if a pure-tone audiometer drifts by a few decibels, “normal hearing” can be misdiagnosed as “mild hearing loss”. Calibrate regularly with couplers/mechanical couplers conforming to the 60318 series (air-conduction with couplers, bone-conduction with the mechanical coupler) and keep calibration records.

4. Engineering Practice and Application 🛠️

  • 🏥 Clinical audiometry: use a pure-tone audiometer conforming to 60645-1, follow the standard masking rules for air- and bone-conduction audiometry, and calibrate periodically plus biological calibration.
  • 👷 Occupational hearing conservation: industry health screening can use equipment to 60645-1/-6; abnormal screening results should be referred for diagnostic audiometry.
  • 🦻 Hearing-aid fitting: tympanometry (60645-5) evaluates middle-ear status, and combined with pure-tone thresholds supports hearing-aid selection and gain setting.
  • 🔬 Research and teaching: short-duration signals (60645-3) are used for acoustic-reflex, adaptation and recruitment studies; set stimulus parameters strictly to the signal time characteristics.
  • 📋 Quality system: audiology facilities should maintain an equipment register and calibration schedule, recording date, method and results of each calibration, and evaluate measurement uncertainty.
💡 Tip for newcomers: the “dB HL” (hearing level) on an audiogram is a level relative to the normal hearing threshold, different from the physical sound pressure level dB SPL. Audiometer output is calibrated to the standard reference zero — do not confuse the two types of decibels when reading results.

5. Frequently Asked Questions ❓

❓ Q: Why must a hospital audiometer comply with 60645-1?
A: Part 1 specifies the frequency, level, distortion and signal requirements of a pure-tone audiometer and gives the calibration reference, ensuring hearing-threshold results are consistent and comparable across devices and institutions — the basis for reliable diagnosis.
❓ Q: What does tympanometry (60645-5) measure?
A: It evaluates middle-ear status by measuring the change in ear-canal acoustic impedance/admittance under controlled pressure, diagnosing middle-ear effusion, tympanic-membrane perforation and ossicular-chain lesions, and measures acoustic-reflex thresholds.
❓ Q: What are short-duration signals (60645-3) used for?
A: Acoustic-reflex threshold measurement, recruitment tests and auditory-adaptation studies, requiring precise control of signal duration, envelope and intensity.
❓ Q: What is the difference between screening and diagnostic audiometers?
A: Screening equipment (e.g. devices related to 60645-6/-7) is for rapid initial screening and may simplify frequency and level ranges; diagnostic audiometers (60645-1) have greater accuracy and function. Abnormal screening results require diagnostic re-testing.
❓ Q: How often should an audiometer be calibrated?
A: Follow the manufacturer’s instructions and usage frequency; usually at least a full calibration (including coupler/mechanical-coupler reference) every year, plus periodic biological calibration with a normal-hearing person.
🔒