⭕ ASME B89.1.6-2002 (Reaffirmed 2012) is the authoritative American National Standard for measuring the plain internal diameters of master rings and ring gages. Prepared by the ASME B89 Dimensional Metrology Committee and approved by ANSI on October 29, 2002, it revises B89.1.6M-1984, extending ring qualification up to 21 in. (533 mm), consolidating information into tables, and relating surface texture to tolerance rather than class. The Standard establishes uniform practices for the measurement of master rings or ring gages using horizontal methods, covering geometric quality requirements, the characteristics of the comparison equipment, environmental conditions, and the means to assure that measurements are made with an acceptable level of accuracy. It is a key reference for ring-gage manufacturing, calibration, and laboratory traceability.
The body of the Standard covers scope, definitions, references, requirements of master rings and ring gages, calibration of an identified diameter, and environment, with three nonmandatory appendices: Appendix A discusses the effects of form and form errors on size (geometry), Appendix B provides measurement uncertainty methodology, and Appendix C offers reference design for ISO cylindrical ring blanks. The Standard explicitly states that it does not include measurement methods for rings below 1 mm (0.040 in.); the method for these very small rings should be agreed upon between the manufacturer/laboratory and the customer prior to manufacture or calibration. Key definitions include bilateral tolerance, circularity (roundness), cosine error, cylindricity, diameter, and dimensional stability.
Section 4 specifies the geometric quality of master rings and ring gages. Table 1 gives surface roughness limits; Table 2 gives limits for roundness, taper, or straightness; and Table 3 gives diameter tolerances by tolerance class and size range. Diameter tolerances are stated as bilateral tolerances — one half of the tabulated tolerance applied plus and minus from the specified size. Surface roughness and form limits loosen moderately as diameter increases and tighten as tolerance class improves. The table below gives representative size-tolerance relationships (the authoritative values are in Table 3 of the Standard).
| Diameter Range | High-Grade Typical Diameter Tolerance (Ref.) | Working-Grade Typical Diameter Tolerance (Ref.) | Typical Surface Roughness Limit |
|---|---|---|---|
| ≤ 1 in (25 mm) | ±0.000 02 in order | ±0.000 2 in order | Ra ≈ 4 μin (0.1 μm) |
| 1–4 in (25–100 mm) | ±0.000 05 in order | ±0.000 5 in order | Ra ≈ 4–8 μin |
| 4–12 in (100–300 mm) | ±0.000 1 in order | ±0.001 in order | Ra ≈ 8 μin (0.2 μm) |
| 12–21 in (300–533 mm) | ±0.000 2 in order | ±0.002 in order | Ra ≈ 8–16 μin |
Section 5 prescribes the calibration of an “identified diameter.” The core of the method is the horizontal measurement approach: the ring is set up on a horizontal measuring instrument, a standard length is reproduced by gage-block combinations (Figure 2 shows typical combination techniques), and the indicator system compares the bore diameter point by point. Figure 1 defines the location of the calibrated diameter — the diameter at a specified axial section and angular position of the bore — so that verification results are physically reproducible. The comparison equipment must have adequate resolution, repeatability, and stable measuring force, and its readings must be traceable to higher-level length standards.
Section 6 specifies environmental conditions: measurements should be taken near the standard reference temperature of 20 °C, with the ring, gage blocks, and equipment fully thermally equilibrated, thermal gradients minimized, and humidity and cleanliness controlled. Appendix A explains from a geometric viewpoint how form errors (roundness, taper, cylindricity, straightness) affect the definition and measurement of “size”: when the measured circle is not ideal, different directions and positions yield different diameter values. Appendix B provides a method for building the uncertainty budget, with components including the reference standard (gage-block) uncertainty, indicator error, probe positioning, temperature and thermal-expansion-coefficient differences, and form-error sampling.