🌡️ ASME B40.200-2008, Thermometers, Direct Reading and Remote Reading, is a consolidated standard issued by the ASME B40 Committee on Pressure and Temperature Instruments. It merges four previously separate thermometer standards into a single volume: B40.3 (Bimetallic Actuated Thermometers), B40.4 (Filled System Thermometers), B40.8 (Liquid-in-Glass Thermometers), and B40.9 (Thermowells for Thermometers and Electrical Temperature Sensors). The document provides harmonized terminology and definitions, dimensions, safety requirements, construction and installation guidance, test procedures, and general recommendations for designers, specifiers, purchasers, and users. It supersedes the SAMA standards RC-6-10-1963 and RC-4-1-1962 and was intended to supersede the military specifications MIL-I-17244, MIL-T-19646, and MIL-T-24272.
The Standard covers thermometers divided by indication method into two broad categories: direct reading types (local indication, such as dial-type bimetallic thermometers and glass thermometers) and remote reading (remote indication) types, in which a sealed filled system transmits the sensing-element signal through a capillary to a distant dial (filled system thermometers). The body of the Standard is organized into chapters on terminology and definitions, general recommendations, safety, selection criteria, test procedures, and references, with nonmandatory appendices providing supplementary information.
Rather than prescribing a single instrument construction, the Standard treats the four product families separately by operating principle, giving manufacturers, users, and inspection bodies a common basis for terminology, accuracy, and testing — which in turn supports interchangeability and comparability among products of the same type.
| Family | Former Standard | Operating Principle | Approx. Range | Typical Applications |
|---|---|---|---|---|
| Bimetallic actuated | ASME B40.3 | Helical bimetallic element deflects with temperature; a pointer moves over a dial scale | −40 °C to 540 °C | Local indication on piping, tanks, furnaces, HVAC |
| Filled system | ASME B40.4 | Sealed gas/vapor/liquid filled system transmits signal through capillary to indicator | −200 °C to 650 °C by fill type | Remote indication in process and storage plants |
| Liquid-in-glass | ASME B40.8 | Mercury or organic liquid expands in a glass bulb; liquid column is read against a scale | −100 °C to 500 °C | Laboratories, field calibration, simple local measurement |
| Thermowells | ASME B40.9 | Protective sleeve isolating the thermometer or RTD from the process | Same as installed instrument | Temperature measurement on pressurized piping and vessels |
The Standard standardizes how thermometers are attached to process piping. Common connection forms include the adjustable-angle connection, where the head can be rotated to the desired viewing direction, and the stem connection, where the sensing stem is inserted directly into the pipe, a thermowell, or a tank nozzle. The figures define the section lengths of the stem — the sensing portion, the threaded connection portion, and the insertion (immersion) portion — so the user can obtain sufficient immersion depth in field installations of varying geometry and minimize conduction errors.
The Standard defines permissible indication error through accuracy grade figures and performance tests. Accuracy is expressed as the conformity of indication to an accepted standard or true value — the difference (error) between the true value and the indicated value, typically stated as a percentage of span or as an absolute value. Higher grades permit smaller error. To verify performance, several test procedures are specified:
| Test | Purpose | Typical Requirement (per Standard figures/tables) |
|---|---|---|
| Accuracy test | Confirm indication error stays within the grade tolerance | Compare against a reference thermometer at at least 5 evenly distributed points |
| Vibration test | Confirm stability under plant mechanical vibration | Table 1 gives amplitudes vs. frequency |
| Load / overrange test | Verify recovery of filled systems and structure after overload | After a specified overrange excursion, indication returns to normal |
| Response test | Characterize time to respond to a temperature step | Record time to reach 63.2% of the step change |