BS EN ISO 11357 — Plastics Differential Scanning Calorimetry (DSC) Standard Series

Plastics — Differential scanning calorimetry (DSC) (identical to ISO 11357)

🌡️ BS EN ISO 11357 is the UK adoption by BSI of the ISO 11357 series covering differential scanning calorimetry (DSC) of plastics. DSC is one of the most widely used thermal analysis methods for polymer materials: by measuring the heat-flow difference between the sample and a reference, it reveals endothermic and exothermic behaviour of the material under controlled heating/cooling, allowing the determination of glass transition temperature, melting and crystallization temperature and enthalpy, crystallinity, specific heat capacity, characteristic reaction-curve temperatures and oxidation stability. The series comprises seven parts and is an essential tool for plastics R&D, quality inspection and failure analysis.

Some parts of this series were sourced from scanned documents; the technical content is compiled from publicly available domain knowledge and the authoritative text remains the official BSI publication.

💡 DSC provides the “thermal fingerprint” of a material from a few milligrams of sample. For polymers, the glass transition reflects molecular-chain mobility, the melting enthalpy reflects crystallinity, and the oxidation induction time reflects antioxidant performance — parameters directly linked to processing and end-use performance.

1. Scope and Overall Framework 📌

BS EN ISO 11357 applies to DSC analysis of polymer materials and their articles, including thermoplastics, thermosets and rubbers. The standard specifies instrument calibration, specimen preparation, temperature programmes and data-processing methods, together with definitions and measurement procedures for the various thermal effects. The series is organised as “general principles (Part 1) + specific determination methods (Parts 2-7)”, ensuring consistency of instrument operation and data processing across methods.

The standard emphasises the significant influence of measurement conditions (heating rate, atmosphere, sample mass) on results, and requires complete recording of test conditions in the report to ensure comparability between laboratories.

2. Series Structure and Technical Highlights 🔧

The table below lists the subject and scope of each part.

PartSubjectScopeEdition
BS EN ISO 11357-1General principlesDSC principles, instrument calibration, terminology2016
BS EN ISO 11357-2Determination of glass transition temperatureGlass transition temperature and step height2014
BS EN ISO 11357-3Determination of temperature and enthalpy of melting and crystallizationMelting and crystallization temperature and enthalpy2013
BS EN ISO 11357-4Determination of specific heat capacitySpecific heat capacity2014
BS EN ISO 11357-5Determination of characteristic reaction-curve temperaturesCharacteristic reaction-curve temperatures2014
BS EN ISO 11357-6Determination of oxidation induction time and temperatureOxidation induction time (OIT) and temperature (OOT)2013
BS EN ISO 11357-7Determination of crystallization kineticsCrystallization kinetics2015
⚠️ The reliability of DSC results depends strongly on instrument calibration and fixed conditions. Temperature and heat flow must be calibrated regularly with reference materials (e.g. indium, zinc, tin); changes in heating rate, sample mass, weighing precision and crucible type significantly affect transition temperatures and enthalpies.

3. Key Technical Requirements ⚙️

  • Instrument calibration: Calibration of the temperature and heat-flow axes with reference materials of known melting point and enthalpy; specified calibration intervals and tolerances.
  • Specimen preparation: Specimen form (granules, film, powder), mass range and pre-treatment (drying, annealing), avoiding interference from moisture and thermal history.
  • Temperature programme: Heating/cooling rate (e.g. 10 K/min), start and end temperatures, atmosphere (nitrogen or air) and flow rate.
  • Data processing: Baseline selection, peak integration and determination of transition temperatures (extrapolated onset, peak temperature, midpoint).
  • Test report: Material identification, instrument program, repeatability and uncertainty, ensuring traceability and comparability.

4. Engineering Practice and Application 🛠️

In the plastics industry, the 11357 series is widely used for material development, incoming inspection and quality management. R&D teams use DSC to determine the glass transition and crystallization behaviour of new formulations and optimise processing windows. Quality departments verify raw-material batch consistency with melting temperature and enthalpy. Failure-analysis engineers use oxidation induction time (OIT) to evaluate whether the antioxidant system has failed and diagnose the causes of product ageing.

Specific heat capacity (Part 4) is an important input for thermal design (e.g. injection-moulding cooling and heat-transfer calculations); crystallization kinetics (Part 7) provides data for modelling isothermal and non-isothermal crystallization and guides moulding process optimisation. DSC data has become standard information on material data sheets.

5. FAQ ❓

❓ Q: Why is the glass transition temperature important?
A: Tg determines the temperature at which a polymer changes from the glassy to the rubbery state, directly affecting the upper service temperature, rigidity, toughness and dimensional stability of the article; it is a core parameter for material selection and process design.
❓ Q: How do melting enthalpy and crystallinity relate?
A: The melting enthalpy of a polymer is proportional to its crystallinity. By comparing the measured melting enthalpy with the theoretical enthalpy of a 100% crystalline polymer, the crystallinity can be estimated, commonly used to evaluate semi-crystalline polymers.
❓ Q: How is oxidation induction time (OIT) applied?
A: OIT is measured at high temperature in an oxygen atmosphere and reflects the effectiveness of the antioxidant system. The longer the OIT, the better the oxidative resistance; it is used to compare formulations and assess the degree of ageing.
🔍 Q: Why measure melting under nitrogen but oxidation under air?
A: Under nitrogen, oxidation is avoided so it does not superimpose on the melting endotherm and distort results; for OIT/OOT measurements the test must run in oxygen or air to evaluate thermal-oxidative stability.
🔒