BS EN ISO 15681 — Water Quality Determination of Orthophosphate and Total Phosphorus by Flow Analysis (FIA and CFA)

Water Quality — Determination of Orthophosphate and Total Phosphorus Contents by Flow Analysis (FIA and CFA) (2004/2009, BS 6068-2.86 / 2.87)

💧 BS EN ISO 15681 is a British Standard series for determining orthophosphate and total phosphorus in water, prepared by ISO/TC 147 (Water quality) with CEN/TC 230 (Water analysis) and published in the UK as BS 6068-2.86 / 2.87 (confirmed 2009). Phosphorus is a key control indicator of eutrophication; accurate determination of orthophosphate and total phosphorus is essential for drinking water, surface water, wastewater monitoring and eutrophication assessment. The series specifies two methods based on flow analysis: Part 1 by flow injection analysis (FIA) and Part 2 by continuous flow analysis (CFA).

💡 Flow analysis performs the reaction in a flowing carrier stream, offering a high degree of automation, low sample and reagent consumption, fast analysis and good repeatability — particularly suitable for routine monitoring of phosphorus in large numbers of water samples.

1. Overview of the Series 📌

Classified under ICS 13.060.50 (Water quality — examination of chemical properties) and published in 2004 with subsequent revisions and confirmation in 2009, both parts belong to the BS 6068 water quality family. The core method is the molybdenum-blue colour reaction: orthophosphate reacts with molybdate in acid medium to form a phosphomolybdenum heteropoly acid which is reduced to molybdenum blue, measured colorimetrically at a defined wavelength; for total phosphorus, a persulphate digestion step first converts organic phosphorus and polyphosphates into orthophosphate.

Standard NumberTitleMethodTypical Use
BS EN ISO 15681-1:2004
(BS 6068-2.86:2004)
Water quality — Determination of orthophosphate and total phosphorus contents by flow analysis — Part 1: Method by flow injection analysis (FIA)Flow injection analysisAutomated determination of large sample batches
BS EN ISO 15681-2:2004
(BS 6068-2.87:2004)
Water quality — Determination of orthophosphate and total phosphorus contents by flow analysis — Part 2: Method by continuous flow analysis (CFA)Continuous flow analysisAutomated determination of large sample batches
⚠️ Phosphorus determination demands a clean environment: phosphorus is ubiquitous in detergents, vessels and reagents. Avoid phosphorus contamination during sampling, storage and laboratory operations. Use dedicated glassware, analytical-reagent-grade chemicals, and rinse sampling equipment thoroughly before use.

2. Structure and Technical Essentials 🔧

FIA and CFA both belong to flow analysis but differ in liquid transport and reaction mode:

  • Flow injection analysis (FIA): a small “plug” of sample is injected into a continuously flowing carrier; while being dispersed and reacting, the sample is transported to the detector and the peak signal is measured at a fixed time. FIA systems are compact and comparatively fast.
  • Continuous flow analysis (CFA): sample and reagent are driven by a peristaltic pump through reaction coils as a continuous stream; air segmentation and mixing accomplish the reaction and the signal is measured under steady-state conditions. CFA is well suited to long-term stable operation with complex matrices.
ComparisonPart 1 — FIAPart 2 — CFA
Sample introductionSample plug injected into carrierContinuous aspiration and segmentation
Reaction modeDispersion and reaction simultaneousSteady-state continuous reaction
System structureCompact and simpleIncludes air-segmentation device
Analysis speedFasterStable; suited to long runs
Detection principleMolybdenum blue colorimetryMolybdenum blue colorimetry

3. Key Technical Requirements: Principle and Reagents 🔬

The Standard prescribes clear requirements for the colour reaction and reagents to ensure accuracy and sensitivity:

  • Molybdenum blue colour reaction: orthophosphate reacts with ammonium molybdate in acid medium to form a phosphomolybdenum complex, which is then reduced (e.g., by ascorbic acid) to a blue species measured at about 880 nm.
  • Total phosphorus digestion: for total phosphorus, the sample is oxidized by persulphate (or digested on-line in the flow system) to convert organic and condensed phosphates to orthophosphate before colour development.
  • Reagent purity: ammonium molybdate, ascorbic acid, persulphate etc. should be analytical grade, and water for reagent preparation should be high-purity to avoid phosphorus contamination.
  • Calibration: prepare a series of standards from potassium dihydrogen phosphate, plot a concentration versus absorbance (or peak height) calibration curve and check it periodically.
✅ Each analytical batch should include blanks, standards, quality-control samples and duplicate samples. Blank values should be below the specified limit; QC results should agree with their nominal values; spike recoveries should fall within the required range to guarantee data reliability.

4. Key Technical Requirements: Instrumentation and Operation 🎯

  • Flow analysis system: peristaltic pump, injection valve, reaction coils, colour development and detection unit, and recording/data-processing system. The system should have stable flow-rate and temperature control.
  • Injection and rinsing: samples and standards are injected alternately; the tubing is flushed with carrier or rinse solution between injections to prevent carry-over.
  • Wavelength selection: a spectrophotometric detector operating at an appropriate wavelength gives sensitivity and selectivity for molybdenum blue.
  • Sample pre-treatment: analyse samples promptly after collection or store per the Standard; filter turbid samples; take care to prevent tubing blockage when suspended solids are present.
🚨 Digestion and reagent preparation involve strong acids and oxidants. Work in a well-ventilated area and wear protective equipment; collect and dispose of organic solvent waste separately, never pour it into drains.

5. Engineering Practice and Applications 🛠️

  • 🔍 Method selection: both FIA and CFA can be used for orthophosphate and total phosphorus; laboratories may choose according to instrument configuration and throughput, and the results of the two methods must be comparable.
  • 📊 Eutrophication monitoring: phosphorus is a major limiting factor for eutrophication in lakes, reservoirs and coastal waters; orthophosphate and total phosphorus data underpin water-quality assessment, load control and the evaluation of remediation measures.
  • 📋 Quality control: build a complete QC framework with blanks, calibration, QC samples and spike recoveries, and take part in inter-laboratory comparisons to ensure mutual recognition of data.
  • 🧪 Typical applications: phosphorus-load monitoring of drinking water, surface water, groundwater, domestic sewage, industrial effluent, agricultural run-off and aquaculture wastewater, plus routine influent/effluent phosphorus checks at wastewater treatment plants.

6. Frequently Asked Questions ❓

❓ Q: What is the difference between orthophosphate and total phosphorus, and how do I choose?
A: Orthophosphate is the dissolved phosphate directly available to biota; total phosphorus includes orthophosphate, condensed phosphates and organic phosphorus. Measure orthophosphate when assessing algal growth potential; measure total phosphorus when evaluating phosphorus load and eutrophication status.
❓ Q: Which method is better, FIA or CFA?
A: Both are internationally recognized valid methods meeting the required accuracy. FIA is simpler and faster; CFA runs stably under steady-state conditions. The choice depends on the laboratory’s equipment, sample throughput and maintenance capability.
❓ Q: Why is digestion needed for total phosphorus?
A: A large part of total phosphorus is organic and condensed phosphate, which does not react directly with molybdate. Persulphate oxidation digestion converts it to orthophosphate, which can then be measured together with the orthophosphate by the molybdenum blue method.
🔍 Q: How do I avoid high blanks caused by phosphorus contamination?
A: Use dedicated acid-washed glassware, prepare reagents with high-purity water and analytical-grade chemicals, thoroughly clean sampling and storage containers, keep the working area clean, and monitor the reagent blank regularly.
🔒