☢️ BS EN ISO 11137 specifies requirements for the radiation sterilization of health care products (primarily medical devices), a core standard in the sterile-assurance system for medical devices. The series has two parts: BS EN ISO 11137-1 covers requirements for development, validation and routine control of the sterilization process, and BS EN ISO 11137-2 covers establishing the sterilization dose. Radiation sterilization uses gamma rays, electron beams or X-rays to irradiate the product and inactivate microorganisms; it penetrates well, can sterilise products in their final packaging, and leaves no residue — widely used for single-use medical devices, dressings and pharmaceutical packaging.
In the UK this series is the responsibility of Technical Committee CH/198 (Sterilization and Associated Equipment and Processes), within ISO/TC 198 “Sterilization of health care products”. BS EN ISO 11137-1:2015 is identical to ISO 11137-1:2006 (including Amd 1:2013), was approved by CEN in June 2015, and supersedes the withdrawn BS EN ISO 11137-1:2006+A1:2013; BS EN ISO 11137-2:2015 is identical to ISO 11137-2:2013, was approved by CEN in June 2015, and supersedes the withdrawn BS EN ISO 11137-2:2013.
| Part | Subject | ICS | ISO Equivalent | Core Content |
|---|---|---|---|---|
| BS EN ISO 11137-1:2015 | Development, validation and routine control | 11.080.01 | ISO 11137-1:2006 (incl. A1:2013) | Requirements and quality management for radiation sterilization processes |
| BS EN ISO 11137-2:2015 | Establishing the sterilization dose | 11.080.01 | ISO 11137-2:2013 | Methods for establishing and verifying the sterilization dose |
BS EN ISO 11137-1 specifies requirements for development, qualification and routine control of radiation sterilization processes, including: product families and grouping; facility and equipment qualification (Installation Qualification IQ, Operational Qualification OQ, Performance Qualification PQ); process definition and dose-mapping; use of dosimeters; routine control and release of sterilization batches; revalidation on process change; and records and documentation.
The Standard requires the process to be built on a quality management system and combined with ISO 14971 risk management: through the SAL 10⁻⁶ target, bioburden monitoring, dose audit and microbiological monitoring, each batch achieves the required sterile-assurance level. It also specifies grouping criteria so that a separate process need not be established for each product.
BS EN ISO 11137-2 specifies the methods for establishing the sterilization dose. The Standard provides two main methods:
After the dose is established, it must be periodically verified by dose audit: testing the bioburden of routine batches to confirm the continued validity of the dose. The Standard gives the specific steps, experimental design, statistical treatment and dose-misuse handling rules for each method.
| Comparison | Method 1 | Method 2 |
|---|---|---|
| Basis | Initial bioburden count | Microbial resistance (D₁₀) testing |
| Experimental effort | Relatively small | Larger (resistance distribution measured) |
| Typical use | Routine products, stable bioburden | Precise dose, highly resistant organisms |
| Verification | Periodic dose audit | Periodic dose audit |
The core of radiation sterilization process control is dosimetry. The Standard requires the use of calibrated dosimeters to determine the minimum and maximum dose distribution in the load configuration, and specifies the routine minimum-dose measurement requirement for batch release. Dosimeter types include alanine/EPR, calorimetric, and radiochromic film, with uncertainty and calibration frequency per the requirements.
Radiation sterilization is widely used for terminal sterilization of single-use syringes, infusion sets, surgical instruments, dressings, implants and pharmaceutical packaging. Manufacturers (of medical products) and contract sterilizers establish and maintain the process per this series, and regulators assess the product’s sterile-assurance compliance. When choosing radiation sterilization, evaluate the radiation resistance of the product and packaging (degradation, discoloration, performance loss) and make an overall decision considering dose and cost.