💧 BS EN ISO 9963 is the British Standard series for the determination of alkalinity in water, prepared by the CEN water quality analytical methods subcommittee (UK counterpart EH/3/2) and identical to the ISO 9963 international standards. Alkalinity is a key indicator of a water body’s buffering capacity, influencing treatment process design and corrosion tendency. The series is divided into two parts: Part 1 specifies the determination of total and composite alkalinity; Part 2 specifies the determination of carbonate alkalinity. Together they form the standard method framework for alkalinity analysis, widely referenced in drinking water, surface water, industrial wastewater and treatment process monitoring.
The series applies to natural water, drinking water, domestic and industrial wastewater, and process waters. The core method is acid-base titration: a standard strong-acid solution (such as hydrochloric acid) is used to titrate the sample, and the content of alkaline species (mainly carbonates, bicarbonates and hydroxides) is determined. Alkalinity is usually expressed as an equivalent of calcium carbonate (CaCO₃) or bicarbonate (HCO₃⁻), in mg/L.
The standard gives clear definitions of “total alkalinity”, “composite alkalinity” and “carbonate alkalinity”: total alkalinity is the total neutralising capacity measured to the methyl orange/mixed-indicator endpoint; composite alkalinity reflects the combined contribution of carbonate and hydroxide; carbonate alkalinity specifically measures the alkalinity contribution of the carbonate form. By measuring these separately, the distribution of carbonate, bicarbonate and hydroxide in the water can be inferred.
| Standard Part | Subject | Measured Quantity | Key Content |
|---|---|---|---|
| BS EN ISO 9963-1:1996 | Total and composite alkalinity | Total alkalinity, composite alkalinity | Principle, reagents, apparatus, sample handling, procedure, result expression |
| BS EN ISO 9963-2:1996 | Carbonate alkalinity | Carbonate alkalinity | Principle, reagents, apparatus, sample handling, procedure, result expression |
Both parts share the same titration principle and sample pre-treatment requirements, but differ in indicator, endpoint and result calculation. Part 2 requires protection against carbon dioxide absorption from air during titration, using a dedicated apparatus to absorb atmospheric CO₂, to ensure accurate carbonate alkalinity measurement.
A standard acid solution of known concentration is added to the sample, progressively neutralising the alkaline species; an indicator changes colour at a specific pH endpoint, and the alkalinity is calculated from the acid consumed. Because carbonate and bicarbonate have two pH transition regions (around pH 8.3 and pH 4.3), carbonate alkalinity and total alkalinity can be obtained from the two endpoints respectively.
The standard specifies hydrochloric acid standard solution, sodium carbonate primary standard, several indicators (methyl orange, phenolphthalein or mixed indicators) and laboratory water complying with ISO 3696. Apparatus includes burettes, conical flasks, magnetic stirrers, an optional pH meter and a carbon-dioxide protection device. All reagents should be analytical grade and the standard solution standardised against a primary standard.
Samples should be analysed as soon as possible after collection; if immediate analysis is impossible, they must be stored refrigerated and analysed within a specified period. Samples with suspended solids or colour require clarification, filtration or appropriate dilution to avoid interference with the endpoint. High-concentration samples may be diluted, while low-concentration samples require a more sensitive indicator or a larger sample volume.
The standard sets out a systematic procedure: transfer a measured volume of sample into a conical flask; add indicator; titrate with the hydrochloric acid standard solution to the endpoint colour change; record the volume consumed; calculate the alkalinity and convert to CaCO₃ (mg/L) or other units. Results are typically expressed as “mg/L CaCO₃” or “mmol/L”; Annex A of the standard provides conversion factors between different units (CaCO₃, bicarbonate, etc.).
To assess method reliability, the standard provides precision data (Table 1) and a comparison of results with other alkalinity methods (Table 2), and requires the test report to include sample identification, sampling information, results, units, precision and the date of determination.