BS EN ISO 07393 — Water Quality Determination of Free Chlorine and Total Chlorine (DPD Titrimetric, DPD Colorimetric and Iodometric Methods)

Water Quality — Determination of Free Chlorine and Total Chlorine: Part 1 / Part 2 / Part 3 (BS EN ISO 7393 series)

💧 BS EN ISO 07393 (ISO 7393) is the method series for determining free chlorine and total chlorine in water, forming the basis for evaluating disinfection effectiveness in drinking water, swimming-pool water, industrial circulating water, and wastewater. This batch covers three parts: Part 1 is the N,N-diethyl-1,4-phenylenediamine (DPD) titrimetric method, Part 2 is the DPD colorimetric method (for routine control), and Part 3 is the iodometric titration method for total chlorine. Developed by ISO/TC 147 (Water quality) and adopted by BSI as a British Standard, it is cross-referenced with Part 2 of the BS 6068 series (BS 6068-2.25 / 2.26 / 2.27).

💡 Free chlorine (free residual chlorine) is a key indicator of water disinfection effectiveness; total chlorine additionally includes chloramines formed with ammonia and organic nitrogen. ISO 7393 uses the DPD colour system to distinguish “free chlorine” from “total chlorine” and provides a complete gradient from precise laboratory titration to rapid field colorimetry, making it one of the most widely used chlorine-determination systems in water quality management.

1. Series Overview and Scope 📌

StandardTitleYear
BS EN ISO 7393-1:2000 (2008)Determination of free chlorine and total chlorine — Titrimetric method using DPD2000 (2008)
BS EN ISO 7393-2:2000 (2008)Colorimetric method using DPD for routine control purposes2000 (2008)
BS EN ISO 7393-3:2000 (2008)Iodometric titration method for the determination of total chlorine2000 (2008)

In the DPD method, DPD reacts with free chlorine to form a red compound, which can be quantified with a spectrophotometer/colorimeter (colorimetric) or titrated with a standard ferrous ammonium sulfate solution until the red colour disappears (titrimetric); after adding potassium iodide, combined chlorine (chloramines, etc.) also participates, so total chlorine is obtained. Part 3 (iodometric titration) uses potassium iodide to displace chlorine and titrates the liberated iodine with sodium thiosulfate; it suits samples with higher total chlorine, such as high-concentration disinfectant solutions and wastewater.

⚠️ Interference and storage: chlorine in water samples is unstable — measure immediately after sampling; if analysis must be delayed, store cold and protected from light and analyse promptly. Oxidising/reducing substances, turbidity, and colour can interfere with DPD colorimetry; use blanks and pretreatment as needed.

2. Series Composition and Technical Points 🔧

2.1 BS EN ISO 7393-1: DPD Titrimetric Method

This part is the DPD titrimetric method, suitable for precise determination of free and total chlorine over a wide range. Add DPD reagent and buffer to the sample; free chlorine turns the DPD red; titrate with ferrous ammonium sulfate standard solution until the colour disappears, and calculate the free chlorine concentration from the consumption; then add potassium iodide so that combined chlorine participates, and continue titrating to the endpoint to obtain total chlorine. The method has simple apparatus and high accuracy, and is the reference method for laboratory arbitration and calibration of field colorimeters.

2.2 BS EN ISO 7393-2: DPD Colorimetric Method

This part is the DPD colorimetric method, suitable for routine control and field/online rapid testing. After the sample reacts with DPD reagent to develop colour, measure the absorbance at the specified wavelength (about 510 nm) with a spectrophotometer or colorimeter, and read the chlorine concentration from a calibration curve or the instrument’s built-in curve; as in Part 1, add potassium iodide to distinguish free and total chlorine. The method is fast and ideal for high-throughput samples and field monitoring.

2.3 BS EN ISO 7393-3: Iodometric Titration Method

This part is the iodometric titration method for total chlorine, suitable for samples with higher chlorine concentration (e.g., disinfectant stock solution, high-chlorine wastewater). In acid, chlorine oxidises potassium iodide to liberate an equivalent amount of iodine; titrate with a sodium thiosulfate standard solution using starch as the indicator until the blue colour disappears, and calculate the total chlorine from the consumption. The method is free from turbidity/colour interference, but the response to chloramines must be handled per the standard.

💡 Method selection: use the DPD titrimetric method (Part 1) for precise laboratory determination and arbitration; the DPD colorimetric method (Part 2) for field and online monitoring; and the iodometric titration method (Part 3) for high-concentration total chlorine samples (e.g., disinfectant stock, high-chlorine wastewater). The three should cross-validate within their applicable ranges.

3. Key Technical Requirements and Test Elements 🎯

  • Sampling and storage: use clean glass or polyethylene bottles, avoid light and bubbles; measure immediately after sampling; if storage is unavoidable, keep cold (e.g., 4 °C) and protected from light, and analyse within the specified time.
  • Reagents and standard solutions: prepare and standardise the DPD reagent, buffer, potassium iodide, ferrous ammonium sulfate, and sodium thiosulfate standard solutions per the specification; store the DPD reagent away from light and cold to prevent oxidation.
  • pH and colour development: DPD colour development occurs in a weakly acidic buffered medium (pH about 6.2–6.5); after adding potassium iodide, the pH and reaction time affect the release of combined chlorine — follow the standard strictly.
  • Instrument and wavelength: the colorimetric method uses a spectrophotometer or colorimeter; set the path length and wavelength per the standard; calibrate the instrument regularly and keep the cells clean and unscratched.
  • Result calculation and reporting: express results in mg/L (as Cl) for free and total chlorine; report the method, sampling and measurement time, and any sample pretreatment.
🚨 Safety: high-concentration chlorine solutions are strongly oxidising and irritating — wear protective goggles and gloves and avoid inhaling chlorine gas; store titrants and DPD reagents per chemical-safety requirements, and collect waste for licensed disposal.

4. Engineering Practice and Applications 🛠️

  • Drinking water and municipal treatment: monitor free and total chlorine in finished water and at the network extremities to ensure disinfection meets the residual-chlorine limits of drinking-water standards such as GB 5749.
  • Swimming-pool and recreational water: routinely test free residual chlorine and combined chlorine (chloramines) to control disinfection and water exchange, ensuring hygienic safety.
  • Industrial circulating water and wastewater: evaluate the dosing effect of chlorine-based biocides in circulating cooling water, paper and food-processing water, and monitor wastewater disinfection and dechlorination.
  • Disinfectant and chlorination-system calibration: calibrate online residual-chlorine analysers by the DPD titrimetric method and verify the dosing of chlorinators/hypochlorite generators.

5. Frequently Asked Questions ❓

❓ Q: What is the difference between free chlorine and total chlorine?
A: Free chlorine (free residual chlorine) is the chlorine with disinfecting activity present as hypochlorous acid/hypochlorite ion; total chlorine also includes chloramines (combined chlorine) formed with ammonia and organic nitrogen. The DPD method measures free chlorine first, then total chlorine after adding potassium iodide; the difference is combined chlorine.
🔍 Q: Why should a water sample not be stored long before chlorine measurement?
A: Chlorine is volatile and oxidising, and continues to react with organic matter and ammonia in the water; light and temperature accelerate its loss. Measure immediately after sampling, or store cold and protected from light and analyse promptly, otherwise results are biased low and unreliable.
❓ Q: Do the DPD colorimetric and titrimetric methods give consistent results?
A: They are based on the same DPD colour development and should agree within their applicable ranges; but colorimetry is affected by turbidity, colour, and path-length error, while titration is more precise. Use colorimetry for rapid field measurement and titration for arbitration when disputed.
🔍 Q: Why is the iodometric titration method suitable for high-chlorine samples?
A: The iodometric method titrates the chlorine-iodine-sodium thiosulfate equivalent relationship; the higher the concentration, the more titrant consumed and the smaller the relative error, and it is free from turbidity/colour interference — hence it suits disinfectant stock and high-chlorine wastewater, whereas the DPD method is better for low-concentration drinking water.
🔒