🛢️ BS EN ISO 6978 specifies methods for determining the mercury content of natural gas, addressing the important engineering problem that gas from certain fields may carry mercury. Although mercury in natural gas is usually present in trace amounts, it forms amalgams with aluminium and can cause stress-corrosion cracking of aluminium heat exchangers, pipe bends, valves and liquefied natural gas (LNG) equipment, and is harmful to personnel. Mercury content is therefore a key parameter in gas quality evaluation and commercial transfer. In the UK the series is implemented under Technical Committee PTI/15, “Natural gas and gas analysis”, as the official British adoption of ISO 6978 and EN ISO 6978, classified under ICS 75.060. The series comprises two parts, using two different sampling sub-methods.
🚨 Mercury is highly toxic and volatile: throughout sampling and analysis, wear protective equipment and work in a well-ventilated area; treat mercury-containing waste and spent adsorption tubes as hazardous waste and never discharge them freely.
1. Scope and General Framework 📌
The series applies to the determination of mercury in natural gas, including elemental mercury and, as applicable, other forms. Because mercury in natural gas is typically at the ng/m³ to µg/m³ level, the standard follows a “sample enrichment first, instrument measurement second” strategy: a measured volume of gas passes through an enrichment medium that captures the mercury, which is then measured by an analytical instrument and divided by the sampled volume to give the concentration.
- BS EN ISO 6978-1:2005 (iodine chemisorption): the gas passes through an iodine-impregnated tube; mercury is captured by chemisorption on iodine.
- BS EN ISO 6978-2:2005 (gold/platinum amalgamation): the gas passes through a gold/platinum alloy tube; mercury forms an amalgam on the alloy surface.
Both sampling methods are used with downstream measuring instruments such as atomic absorption spectrometry (AAS), together forming a complete mercury determination procedure. The series supersedes the earlier UK BS 7678:1993 and complements ISO 6978-3 (other methods) to form a complete system.
💡 The two sampling options each have advantages: the iodine chemisorption method tolerates a wider range of mercury forms; the gold/platinum amalgamation method has high capture efficiency for elemental mercury and low blank. Laboratories choose by gas quality (sulfur, condensate content, etc.).
2. Series Structure and Technical Highlights 🔢
| Standard | Title | Sampling principle | Typical companion measurement |
| BS EN ISO 6978-1:2005 | Natural gas — Determination of mercury — Part 1: Sampling of mercury by chemisorption on iodine | Iodine chemisorption enrichment | Atomic absorption spectrometry (AAS) and similar trace-mercury analysis |
| BS EN ISO 6978-2:2005 | Part 2: Sampling of mercury by amalgamation on gold/platinum alloy | Gold/platinum amalgamation enrichment | Thermal desorption—AAS mercury analysers |
The two parts follow the same principles for sampling-system architecture (sampling line, flow control, adsorption tube, gas volume metering), sample protection and blank control; the difference is mainly in the enrichment medium and the corresponding desorption/pre-treatment before measurement.
3. Sampling System and Key Steps 🎯
The basic flow of mercury determination includes:
- Sampling-system setup: draw gas from the sample point, pass it through pressure reduction and flow control, then through the adsorption tube, metering the volume with a wet gas meter or mass flow meter; keep the system short and heated to prevent condensation.
- Enrichment capture: mercury in the gas is captured on iodine (Part 1) or gold/platinum alloy (Part 2); the sampled volume is set from the expected content and adsorption capacity to avoid breakthrough.
- Sample preservation and transport: seal the enriched tube to prevent mercury loss and contamination, and analyse promptly.
- Desorption and measurement: heat or chemically treat the tube to release the mercury into the AAS or similar instrument; correct with blank tubes and standard solutions.
- Result calculation: divide the measured mercury by the sampled gas volume (converted to the specified reference conditions) to obtain the content, usually expressed in µg/m³.
⚠️ The most common error sources in sampling are “mercury breakthrough” and “pre-tube adsorption”: too high a flow rate or volume can cause breakthrough, while adsorption on the inner wall, valves and fittings of the line causes a low result. Perform breakthrough tests per the standard and thoroughly passivate/clean the sampling line.
4. Quality Control and Result Expression 📊
To ensure the reliability of trace measurement, pay attention to:
- Blank control: carry blank tubes with each batch and subtract the blank contribution of sampling and analysis.
- Recovery verification: verify total sampling and analysis recovery using a standard mercury source or by spiking.
- Duplicate sampling: take at least two parallel samples at the same point and check consistency.
- Units and reference conditions: express results in µg/m³ and state the reference temperature/pressure for volume conversion (e.g. 0 °C, 101.325 kPa or 20 °C, 101.325 kPa).
✅ Recommended practice: before gas commercial transfer and equipment selection (especially LNG and aluminium exchangers), carry out systematic mercury testing to the ISO 6978 series, and combine with composition standards such as ISO 6974 to form a complete gas-analysis report.
5. Engineering Practice 🛠️
- 🔍 Gas quality evaluation: assess mercury content before field development and pipeline transport to identify high-risk sources.
- 📊 Equipment protection: judge whether a mercury-removal unit (e.g. activated carbon, sulfur-impregnated carbon) is needed to prevent amalgam corrosion of aluminium exchangers.
- 📋 Commercial transfer: specify the mercury limit and test method in gas purchase contracts, protecting both buyer and seller.
- 🔧 Occupational health and environment: monitor mercury exposure in production and processing, and implement safety protection and waste management.
6. Frequently Asked Questions ❓
❓ Q: Why is mercury in natural gas dangerous?
A: Mercury forms amalgams with aluminium, causing stress-corrosion cracking of aluminium heat exchangers and equipment; mercury vapour is also toxic to humans, so special protection is needed when handling LNG and gas plants.
❓ Q: Can Part 1 and Part 2 results be used interchangeably?
A: Within the validated range yes, but the applicable gas quality (hydrogen sulfide, condensate, etc.) and detection limits differ slightly. Always state which part was used.
❓ Q: What is the lower detection limit?
A: By increasing sample volume and optimising measurement, mercury at the ng/m³ level can typically be detected; the exact value depends on the instrument and method validation.
❓ Q: Does sulfur-containing gas affect the measurement?
A: Yes. Sulfur compounds such as hydrogen sulfide may interfere with adsorption and measurement; use the standard’s recommended pre-treatment or the more tolerant sampling method, and verify by comparison if necessary.
❓ Q: Does the series measure only elemental mercury?
A: The standard targets mercury in natural gas, with elemental mercury the main focus; in practice the sampling and analysis conditions capture measurable forms, and the exact coverage depends on the standard text and laboratory validation.