BS EN 62586 — Power Quality Measurement in Power Supply Systems (Power Quality Instruments)

Power Quality Instruments — Requirements, Functional Tests, and Uncertainty

⚡ The BS EN 62586 series is the British Standard adoption of the IEC 62586 series, specifying the requirements, functional tests, and measurement uncertainty for power quality instruments (PQI) used to measure power quality in supply systems. With the rapid growth of power electronics, renewable generation, and EV charging loads, grid harmonics, voltage sags/swells, flicker, frequency deviation, and three-phase unbalance have become increasingly prominent — accurate and comparable power quality measurement is now a foundation of grid operation and energy management.

💡 The series is maintained by UK Technical Committee PEL/85 (Measuring equipment for electrical and electromagnetic quantities), exactly mirroring IEC/TC 85. Adopting this series ensures comparable and consistent measurement results across different manufacturers’ instruments.

1. Series Structure and Standard Framework 📌

The BS EN 62586 series currently comprises two parts:

Standard No.TitleMain PurposeYear
BS EN 62586-1Power quality measurement — Part 1: Power quality instruments (PQI)Terms, performance, construction and safety requirements2014
BS EN 62586-2Power quality measurement — Part 2: Functional tests and uncertainty requirementsFunctional test methods and measurement uncertainty limits2014

Part 2 also incorporates the IEC November 2014 corrigendum, with amended text flagged in the document. The series covers instruments measuring voltage, current, frequency, harmonics, interharmonics, flicker, and sag/swell events — usable for fixed online monitoring or as portable instruments for field diagnosis.

⚠️ A PQI differs from an ordinary energy meter: it must capture and record transient events (e.g., a sag lasting tens of milliseconds) and its measurement algorithms must comply with the standard; otherwise the data cannot be compared with standard limits.

2. Part 1: Requirements for Power Quality Instruments 🔧

BS EN 62586-1:2014 specifies the basic requirements for PQI-class instruments. Main content includes:

  • Terms and classification: defines power quality instruments and measurement classes (Class A / Class S / Class B, etc.) with their applications, harmonising harmonic and flicker terminology.
  • Measurement functions: specifies the quantities the instrument must measure, including supply voltage characteristics (magnitude, frequency, unbalance), harmonics and interharmonics, flicker (short-term Pst and long-term Plt), and sag/swell/interruption event recording.
  • Construction and interfaces: specifies electrical, communication, display, and human-machine interfaces for configuration and data retrieval.
  • Safety and environmental requirements: specifies electrical safety in the grid environment (overvoltage category, pollution degree), insulation, and EMC requirements.
  • Power supply: clarifies that the instrument may be powered from the measured system or independently, and requires robustness over the operating voltage/frequency range.
✅ Part 1 fixes “what to measure and how” — it establishes the functional framework for power quality parameter measurement, providing the object for the performance verification of Part 2.

3. Part 2: Functional Tests and Measurement Uncertainty 📊

BS EN 62586-2:2014 is the core performance standard of the series, specifying verification methods and uncertainty limits:

  • Functional tests: inject specified test signals from a standard source to verify measurement accuracy for voltage, current, frequency, harmonics, flicker, and event parameters.
  • Measurement uncertainty requirements: gives allowable uncertainty limits for each measured parameter — voltage magnitude, frequency, harmonic magnitude and phase, flicker, etc. must meet specified maximum permissible errors.
  • Test conditions: specifies environmental (temperature, humidity), supply, and data-processing (e.g., 10/12-cycle windowing) conditions for functional tests.
  • Conformity assessment: used for type and surveillance tests to confirm the instrument meets uncertainty requirements across its range — the basis for product conformity decisions.
🚨 The value of power quality data depends on measurement uncertainty: if the instrument’s own error or algorithm is non-compliant, an over-limit harmonic conclusion is unreliable and may cause misjudgement or missed detection, leading to wrong mitigation investments.

4. Key Technical Requirements 📏

ParameterMeasurement PointEngineering Focus
Supply voltage magnitudeRMS value and variation over timeAssesses voltage deviation and end-of-line quality
Frequency deviationDeviation of fundamental frequency from nominalReflects system frequency stability; alarm when out of limits
Harmonics and interharmonicsIndividual harmonic magnitude/phase, THDAssesses power-electronics pollution; compare with IEC 61000 limits
Flicker (Pst / Plt)Short- and long-term flicker severityMeasures lighting flicker and voltage fluctuation effects
Sag / swell / interruptionEvent depth, duration, and time of occurrenceDiagnoses sensitive-load tripping and production interruption
Three-phase unbalanceNegative/zero-sequence components and degreeAssesses the effect of unbalanced single-phase loads on transformers
⚠️ Measurement class selection is critical: Class A instruments are used where rigorous, traceable measurement is required (e.g., arbitration of disputes, regulatory conformity); Class S is mainly for statistical surveys; Class B is only for general reference monitoring and must not be used for critical judgement.

5. Engineering Practice 🛠️

  • Grid-side monitoring: install online PQIs at substations and distribution nodes to continuously record voltage quality and alert on limit violations.
  • User-side diagnosis: run targeted sag/flicker measurements for sensitive industrial customers to identify power quality issues affecting equipment.
  • Mitigation evaluation: compare before/after measurements around harmonic filters (active/passive) to quantify performance.
  • Renewable interconnection: monitor power quality at PV, wind, and storage grid connection points to support interconnection assessment and operation.
  • Data management: unify event definitions and statistical methods per the standard to build a comparable power quality database.

6. Frequently Asked Questions ❓

❓ Q: What is the difference between Class A, Class S, and Class B instruments?
A: Class A has the strictest uncertainty requirements and is used for arbitration and conformity judgement; Class S is for statistical surveys; Class B is for general reference monitoring only. Choose the class by measurement purpose.
❓ Q: Does power quality measurement require calibration traceability?
A: Yes. PQIs should be traceable to national standards; the functional tests of Part 2 verify uncertainty conformance, and type tests plus periodic calibration are part of a quality management system.
❓ Q: What is the relationship between this series and IEC 61000-4-30?
A: IEC 61000-4-30 is the measurement-method standard defining algorithms and time aggregation for power quality parameters; BS EN 62586 targets the instrument itself. They are used together.
❓ Q: How do I judge whether a voltage sag caused sensitive equipment to trip?
A: Use a Class A or Class S instrument to record the sag depth, time, and duration, then compare with the equipment voltage-tolerance curve (e.g., ITIC curve) to see if it enters the sensitive region.
🔒