📱 BS EN 62209 is the British Standard implementation of IEC 62209, maintained by BSI Technical Committee GEL/106 (Human exposure to low-frequency and high-frequency electromagnetic radiation). The technical content is identical to IEC 62209. The series specifies measurement procedures for the specific absorption rate (SAR) of human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices (such as mobile phones, tablets, smart watches and two-way radios). It is the core method standard for SAR certification and regulatory conformity assessment of wireless devices worldwide. The series consists of two parts: Part 1 covers devices used next to the ear (300 MHz to 6 GHz), and Part 2 covers devices used in close proximity to the body (30 MHz to 6 GHz).
💡 SAR is the physical quantity that measures the rate at which electromagnetic energy is absorbed in human tissue, expressed in W/kg. Public exposure limits set by regulators (such as Europe, the US FCC and the international ICNIRP) are typically 2 W/kg averaged over 10 g, or 1.6 W/kg averaged over 1 g. BS EN 62209 provides the standard method for “how to measure the SAR value and compare it with the limits” — an essential step in the pre-market evaluation of wireless devices.
1. Overview of the Standard Series 📌
The BS EN 62209 series is maintained by BSI Technical Committee GEL/106 (Human exposure to low- and high-frequency electromagnetic radiation), with UK experts participating in the work of IEC TC 106 (Electromagnetic fields and human exposure). Part 1 (2016 edition) supersedes the 2006 edition and applies to devices used next to the ear; Part 2 (2010 edition) applies to devices used in close proximity to the body. The frequency ranges of the two parts differ: Part 1 covers 300 MHz to 6 GHz, and Part 2 covers 30 MHz to 6 GHz. As stated in the standard, compliance with a British Standard does not confer immunity from legal obligations, and users remain responsible for correct application.
The series is used together with exposure-limit documents such as the ICNIRP guidelines and IEEE C95.1: the standard specifies the “measurement method”, while the limit documents specify the “permissible values”. Only the two together enable conformity assessment of wireless devices. It is worth noting that the 2016 edition of Part 1 places particular emphasis on measuring SAR across multiple typical use postures (such as different angles next to the ear) and multiple operating modes (different bands and power levels), and taking the maximum value.
2. Structure and Scope of the Series 🧩
The BS EN 62209 files covered by this article comprise two parts:
| Standard number | Year | Title | Scope |
| BS EN 62209-1:2016 | 2016 | Part 1: Devices used next to the ear | SAR measurement of hand-held wireless communication devices (principally mobile phones) in the ear position, frequency range 300 MHz to 6 GHz |
| BS EN 62209-2:2010 | 2010 | Part 2: Devices used in close proximity to the human body | SAR measurement of hand-held and body-worn devices (mobile phones, tablets, wearables, two-way radios) in positions close to the body, frequency range 30 MHz to 6 GHz |
3. Key Technical Requirements 🔧
3.1 Measurement System and Human Phantom
SAR measurement relies on a precision automated test system whose core elements include:
- Human-equivalent phantom: a head or torso model filled with tissue-equivalent liquid whose permittivity and conductivity are configured by frequency to simulate the electromagnetic properties of human tissue.
- Field probe: a three-axis, isotropic electric-field probe that scans automatically inside the liquid, measures the electric-field strength and converts it to SAR.
- Robotic positioning system: moves the probe along a predefined grid to measure the three-dimensional SAR distribution.
- Measurement instruments and fixtures: power meters, signal sources, fixtures and positioning devices that ensure the device under test transmits in the standard-specified position and posture.
3.2 Measurement Procedure and Evaluation
- Device posture and configuration: set the relative position of the device and phantom as specified (next to the ear, close to the body, at different distances), and select typical operating bands and maximum transmit power.
- Peak SAR localisation: first perform a coarse scan to locate the peak region, then a fine scan to obtain the peak SAR (averaged over 1 g or 10 g).
- Maximum over postures and bands: measure the device in several typical postures and operating modes separately, and use the maximum value as the evaluation result.
- Uncertainty evaluation: evaluate the contributions of probe calibration, liquid parameters, positioning errors and other sources, and report the expanded uncertainty.
⚠️ The accuracy of SAR measurement depends strongly on the tissue-equivalent liquid parameters. The permittivity and conductivity of the liquid vary with frequency, must be measured and calibrated after preparation, and drift as the temperature changes during testing. Common error sources include liquid parameter deviation, probe-calibration failure and positioning errors, which must be controlled through system validation and intermediate checks.
4. Test Implementation and Conformity Assessment 🧪
SAR testing is usually performed in an accredited third-party laboratory. The implementation flow includes: system validation (using a standard reference dipole to produce a known SAR field inside the phantom to verify system accuracy) → configuration and positioning of the device under test → coarse/fine scan → peak SAR calculation → uncertainty evaluation → report. For conformity assessment, the measured peak SAR is compared with the applicable limit (for example 2 W/kg averaged over 10 g under European EN 50360/EN 50566, or 1.6 W/kg averaged over 1 g under US FCC rules). Because the measurement system has uncertainty, the usual practice is that the measured value plus the expanded uncertainty must still remain below the limit for a pass decision.
✅ Practical suggestion: during the design phase, wireless-device manufacturers should estimate the peak SAR location by simulation (for example FDTD modelling) to guide antenna layout and shielding design. During testing, give priority to covering the “worst-case posture” (normally the antenna closest to the body, maximum power and the lowest/highest bands) to reduce the number of tests while keeping compliance margin.
5. Engineering Practice 🛠️
- 🔍 Selection and regulatory check: when purchasing wireless devices, verify that the SAR test report was produced in accordance with BS EN 62209 (or the corresponding IEC 62209) methods and compare with the exposure limits of the target market (EU, US, China, etc.).
- 🔧 Design optimisation: keeping the antenna away from the body, adopting low-SAR antenna designs and properly setting transmit power and duty cycle are the main ways to reduce SAR.
- 📊 Multi-mode check: devices such as mobile phones support multiple wireless technologies (2G/3G/4G/5G, Wi-Fi, Bluetooth) and multiple bands; SAR testing must cover all typical transmit modes and take the maximum.
- 📋 Documentation and declaration: include the SAR test report and declaration of conformity in the product technical file as evidence for market surveillance and consumer information disclosure.
6. Frequently Asked Questions ❓
❓ Q: Why are there two averaging masses, 1 g and 10 g, for SAR limits?
A: It depends on which exposure-limit system is used. The US FCC uses 1 g averaging with a limit of 1.6 W/kg; Europe (ICNIRP/EN) uses 10 g averaging with a limit of 2 W/kg. Because the averaging mass differs, the SAR values measured for the same device under the two systems are not directly comparable.
❓ Q: Does BS EN 62209 specify the SAR limits?
A: No. BS EN 62209 specifies the “measurement procedure”, that is, how to obtain the SAR value; the limits are set by ICNIRP, IEEE C95.1 and national regulations (such as EU directives and FCC rules). The two must be used together to complete a conformity decision.
❓ Q: How do Part 1 and Part 2 divide the work?
A: Part 1 addresses “devices used next to the ear” (mainly mobile phones), while Part 2 addresses “devices used close to the body” (pocket, chest, armband, wearables). The same mobile phone may need to be measured under both parts depending on its use scenarios; in terms of frequency, Part 2 has a lower limit (30 MHz).
🔍 Q: Why must the uncertainty be added to the measured value before a pass decision?
A: The SAR measurement system has systematic and random errors (probe, liquid, positioning, etc.). If the decision were based on the measured value alone, a measured value below the limit could still correspond to a “true value” above the limit. The standard practice is that the measured value plus the expanded uncertainty must still be below the limit, to protect public safety and reduce the risk of wrong decisions.