🧪 BS EN 60746 is the British Standards Institution (BSI) implementation of the European Standard EN 60746 (derived from IEC 60746) covering the expression of performance of electrochemical analyzers. Electrochemical analyzers (e.g., pH meters, oxidation-reduction potential meters, conductivity meters, dissolved-oxygen analyzers) are widely used water-quality instruments in industrial processes and laboratories. This series unifies the method of expressing performance and the rules for evaluating various electrochemical analyzers, making technical specifications from different manufacturers comparable and providing a common basis for selection, acceptance and metrology management. The UK participation is entrusted to Subcommittee GEL/65/4 (Process instruments for gas and liquid analysis) under Technical Committee GEL/65 (Industrial-process measurement and control).
💡 The core value of this series is standardised “expression of performance”: it unifies the definitions and test methods of metrological characteristics (e.g., error of indication, repeatability, response time), avoiding misleading selection from vendors’ self-declared figures.
1. Series Overview and Scope 📌
The British Standards in this batch cover three parts:
| Standard | Part Content | Corresponding IEC |
| BS EN 60746-1:2003 | Part 1: General (general rules for expression of performance) | IEC 60746-1:2003 |
| BS EN 60746-2:2003 | Part 2: pH value measurement | IEC 60746-2:2003 (with corrigendum July 2003) |
| BS EN 60746-3:2002 | Part 3: Electrolytic conductivity | IEC 60746-3:2002 (with corrigendum January 2003) |
All three parts supersede their earlier UK standards (e.g., BS 6438-1:1984, BS 6438-2:1984, BS 6438-3:1986). Parts 2 and 3 are used in conjunction with the general rules of Part 1. The BS EN 60746 series also covers other electrochemical analyzer parts (e.g., dissolved oxygen, electrolytic water quality).
⚠️ The performance of a given electrochemical analyzer depends heavily on the electrode, temperature compensation, calibration and sample conditions; expression of performance must state the test conditions (temperature, pH buffer, conductivity standard solution, etc.), otherwise the figures cannot be reproduced or compared.
2. General Rules for Expression of Performance (Part 1) 📐
BS EN 60746-1 specifies general rules for expressing performance of electrochemical analyzers:
- Metrological characteristics: measuring range, error of indication (accuracy), repeatability, stability, zero drift, response time, temperature influence, etc.;
- Environmental characteristics: influence of temperature, humidity, supply and electromagnetic compatibility;
- Unified terminology: unified definitions of “sensitivity, resolution, dead band, hysteresis”;
- Test and evaluation: performance test methods and evaluation procedures to ensure credible figures.
The general rules build a complete “definition—test—evaluation” framework; each part (pH, conductivity, ORP, etc.) adds provisions specific to the measured quantity.
3. pH Value Measurement (Part 2) ⚖️
BS EN 60746-2 specifies requirements for expressing the performance of pH meters:
- Measurement principle: a potential measurement system of glass electrode and reference electrode based on the Nernst equation;
- Metrological characteristics: pH range, accuracy, repeatability, slope error, input-impedance requirements;
- Temperature compensation: effect of temperature on pH measurement and automatic/manual compensation requirements;
- Test methods: calibration and verification using standard buffers, evaluating error of indication, etc.
pH meters are the most common electrochemical analyzers in laboratories and process control, widely used in water quality, chemical, pharmaceutical, food and environmental monitoring. The standard unifies the expression of performance of the instrument and electrode system.
💡 pH measurement requires two- or multi-point calibration with certified standard buffers, and attention to electrode activation, cleaning and storage; temperature compensation and electrode ageing are key factors affecting long-term accuracy.
4. Electrolytic Conductivity (Part 3) ⚡
BS EN 60746-3 specifies requirements for expressing the performance of conductivity meters:
- Measurement principle: measuring the conductance (or resistance) of the solution between electrodes, related to ion concentration;
- Metrological characteristics: conductivity range, accuracy, repeatability, temperature coefficient and compensation, cell constant;
- Electrode system: electrode type (two-electrode/four-electrode), determination and verification of cell constant;
- Test methods: verification with standard conductivity solutions, evaluating error of indication and repeatability.
Conductivity meters are widely used for pure/ultra-pure water monitoring, boiler-water treatment, reverse-osmosis systems and process water-quality control. The temperature dependence of conductivity (about 2%/°C) must be compensated during measurement.
🚨 Conductivity measurement is extremely temperature-sensitive (temperature coefficient about 2%/°C). When comparing or accepting instruments, the reference temperature (typically 25 °C) and compensation method must be stated, otherwise readings under different conditions cannot be compared directly.
5. Engineering Application and Practice 🛠️
- 🔍 Selection: choose pH/ORP/conductivity analyzers by medium and range, and verify accuracy, repeatability, response time and protection degree.
- 🔧 Installation and calibration: install electrodes/sensors correctly, calibrate periodically with standard solutions, and maintain electrode condition.
- 📊 Temperature compensation: enable appropriate temperature compensation (automatic preferred) to ensure accuracy.
- 📋 Metrology management: calibrate at set intervals, keep calibration records and a traceability chain to ensure credible results.
In an industrial installation, electrochemical analyzers are usually part of a sample-handling system, so the performance of the analyzer must be considered together with the sampling, conditioning and calibration functions. For pH measurement, the placement of the electrode, the flow rate of the sample and the temperature gradient across the electrode affect the result, and the instrument must therefore be installed and calibrated according to the standard’s recommendations. For conductivity measurement, the cell constant must be verified with a standard solution at the reference temperature, and automatic temperature compensation should be enabled to report values at 25 °C. The standard provides a common basis for these procedures, ensuring that measurements taken by different installations are comparable and that the performance claims of the instruments are meaningful.
6. Frequently Asked Questions ❓
❓ Q: What instruments does BS EN 60746 cover?
A: Electrochemical analyzers such as pH meters, oxidation-reduction potential (ORP) meters, conductivity meters and dissolved-oxygen analyzers; this batch covers pH, conductivity and general rules.
❓ Q: Why is standardisation of performance expression needed?
A: It unifies definitions and test methods of metrological characteristics, making technical specifications from different manufacturers comparable, avoiding selection errors and aiding acceptance and metrology management.
❓ Q: What should be noted when calibrating a pH meter?
A: Use certified standard buffers for two- or multi-point calibration, pay attention to electrode activation, cleaning and storage, and enable temperature compensation.
🔍 Q: Why must temperature be considered in conductivity measurement?
A: Solution conductivity changes significantly with temperature (about 2%/°C); it must be compensated to a reference temperature (25 °C) before comparison and acceptance.
❓ Q: How is accurate electrochemical analyzer measurement ensured?
A: Correct selection and installation, periodic calibration, proper temperature compensation, periodic verification and a traceability record.