BS EN 60318 — Electroacoustics Simulators of Human Head and Ear Standard Series

Electroacoustics — Simulators of Human Head and Ear(IEC 60318 / EN 60318 Series)

👂 BS EN 60318 is the British Standard series that adopts the European Standard EN 60318 (identical to the IEC 60318 series) for simulators of the human head and ear. By specifying standardised ear simulators, occluded-ear simulators, acoustic couplers and mechanical couplers, it provides a “standard ear” for the calibration and measurement of headphones, hearing aids, ear inserts and audiometers — making electroacoustic measurements comparable across laboratories and products. Batch BSB053 covers five documents: 60318-1, -3, -4, -5 and -6.

💡 Why is an “ear simulator” needed? Because the acoustic load of a headphone is the real human ear, and real ears differ from person to person. The 60318 series uses artificial ear canals and couplers that match the statistically average human ear impedance, turning “person-to-person variation” into a repeatable, calibratable standard measurement condition.

1. Scope and General Framework 📌

The 60318 series belongs to the field of electroacoustics, prepared by IEC TC 29, with UK participation through committee EPL/29. Each part specifies an ear simulation device for a different purpose:

  • 60318-1 (2009): ear simulator for the measurement of circumaural and supra-aural earphones. The frequency range is extended to 16 kHz, and the acoustical transfer impedance, including tolerances, is revised with a measurement method.
  • 60318-3 (2015): acoustic coupler for the calibration of supra-aural earphones used in audiometry — the basic device for routine audiometer calibration.
  • 60318-4 (2010): occluded-ear simulator for the measurement of earphones coupled to the ear by means of ear inserts, usable from 100 Hz to 16 kHz.
  • 60318-5 (2006): 2 cm³ acoustic coupler for the calibration of supra-aural earphones used in audiometry.
  • 60318-6 (2008): mechanical coupler for the measurement of bone-conduction earphones.
⚠️ Documents 60318-5 and -6 in this batch are scanned versions, written from public-domain knowledge: the 2 cm³ coupler is the classic calibration cavity for audiometric earphones, and the bone-conduction mechanical coupler simulates skull transmission with a standard mechanical impedance. The technical points are summarised from well-established practice; specific values are governed by the official BSI text.

2. Series Structure and Sub-standard Overview 🗂️

StandardSubjectEditionTypical use
BS EN 60318-1:2009Ear simulator for circumaural/supra-aural earphones2009, replaces 1998 & 60318-2:1998Standard acoustic load for headphone response, sensitivity and distortion
BS EN 60318-3:2015Acoustic coupler for supra-aural audiometric earphones2015, replaces 1998Acoustic calibration of audiometer air-conduction earphones
BS EN 60318-4:2010Occluded-ear simulator for ear-insert coupling2010, replaces BS 6310:1982Insert earphones, hearing aids and in-the-ear devices
BS EN 60318-5:20062 cm³ acoustic coupler2006Simple calibration of audiometric and insert earphones (scanned)
BS EN 60318-6:2008Mechanical coupler for bone-conduction earphones2008Calibration of bone-conduction audiometers and hearing aids (scanned)

3. Key Technical Requirements 🔧

3.1 Ear simulator (60318-1)

Part 1 specifies the ear simulator used for measuring supra-aural and circumaural earphones. The 2009 edition extends the usable frequency range to 16 kHz, revises the specification for the acoustical transfer impedance, including tolerances, and adds a method for measuring it together with expanded measurement uncertainties. The ear simulator presents an acoustic impedance at its entrance plane that matches the statistically average human ear, so frequency response measured with different headphones reflects the real in-ear sound pressure.

3.2 Audiometric acoustic couplers (60318-3, -5)

Part 3 specifies the acoustic coupler for calibrating supra-aural audiometric earphones: the coupler has defined geometric and acoustic properties that convert the headphone sound pressure into a repeatable sound pressure level, used to set the output sound-level reference of an audiometer. The 2015 edition mainly updated reference signal figures and definitions. The 2 cm³ coupler (60318-5) is a simpler, longer-established classic cavity, widely used for routine audiometer calibration.

3.3 Occluded-ear simulator (60318-4)

For insert earphones and hearing aids, the ear canal is occluded in real use, so an occluded-ear simulator is needed: it simulates the acoustic load of an ear canal sealed by an ear insert. The 2010 edition extends the usable frequency range to 100 Hz – 16 000 Hz and adds values of maximum permitted expanded uncertainties to all tolerances, improving cross-laboratory consistency.

3.4 Bone-conduction mechanical coupler (60318-6)

Bone-conduction earphones transmit sound to the cochlea through skull vibration. Part 6 specifies a mechanical coupler: it simulates the skull with a standardised mechanical impedance and, under a defined contact force and area, measures the output force level or equivalent sound pressure level of the bone-conduction earphone — the basis for bone-conduction audiometry and hearing aid verification.

🚨 Calibration chain integrity is critical: the sound-level accuracy of an audiometer directly affects hearing diagnosis. Air-conduction earphones should be calibrated with the 60318-3/-5 couplers, and bone-conduction earphones with the 60318-6 mechanical coupler; the two must not be interchanged, or systematic measurement error results.

4. Engineering Practice and Application 🛠️

  • 🎧 Headphone development and QC: rated frequency response and sensitivity should be measured on the 60318-1 (circumaural/supra-aural) or 60318-4 (insert) simulator, and the simulator model should be stated.
  • 🏥 Audiometer calibration: audiology facilities should periodically calibrate air-conduction earphone levels with the 60318-3/-5 couplers and bone-conduction earphones with the 60318-6 coupler, keeping calibration records and uncertainty evaluation.
  • 🦻 Hearing aid fitting: the difference between 2 cc coupler (60318-5) and real-ear measurements must be understood through coupler-to-real-ear correction values; do not mix readings directly.
  • 📏 Uncertainty management: the 60318 parts generally give measurement uncertainty requirements; laboratories should evaluate uncertainty and take part in periodic inter-laboratory comparisons.
💡 One-line distinction: 60318-1/-4 are “simulated ears” for product development and in-ear-level measurement; 60318-3/-5 are “couplers” for audiometer calibration; 60318-6 is a “mechanical ear” for bone-conduction measurement. Choosing the right device is the prerequisite for valid data.

5. Frequently Asked Questions ❓

❓ Q: What is the difference between an ear simulator and an acoustic coupler?
A: An ear simulator (60318-1/-4) mimics the acoustic impedance of the real ear for development and more realistic headphone measurement; an acoustic coupler (60318-3/-5) is a geometrically and acoustically simple, stable cavity used mainly for routine audiometer calibration.
❓ Q: Why use 60318-3 for supra-aural earphones but 60318-4 for insert earphones?
A: A supra-aural earphone rests on the pinna, approximating a coupler load; an insert earphone seals the ear canal and requires an occluded-ear simulator. The load characteristics differ, so the corresponding device must be used.
❓ Q: Does the 60318-6 mechanical coupler measure “volume”?
A: Bone-conduction measurement yields an output force level or equivalent sound pressure level under a standard mechanical load, used to calibrate the bone-conduction output of an audiometer; it is a different physical quantity from air-conduction sound pressure level and cannot be compared numerically.
❓ Q: How should the stated sensitivity of a headphone be understood?
A: Sensitivity (often dB SPL/1 mW or dB SPL/V) is the output sound pressure level measured on a specified simulator or coupler; the manufacturer must state the coupler and measurement conditions for the value to be meaningful.
❓ Q: Why does measurement uncertainty matter?
A: The same headphone can give different results in different laboratories. The 60318 parts specify expanded uncertainty and calibration methods to help laboratories assess reproducibility and set reasonable error limits for hearing diagnosis.
🔒