📡 BS EN 50147 is the CENELEC standard series for anechoic chambers (also known as electromagnetic shielded rooms), defining methods for measuring the shielding performance of anechoic chambers used for electromagnetic compatibility (EMC) testing. The series is centred on BS EN 50147-1, which gives the general method, test layout, frequency range, instrumentation, and data-handling requirements for measuring shield attenuation — a foundation document for the construction, acceptance, in-service verification, and EMC compliance testing of EMC laboratories.
💡 The “shield attenuation” of an anechoic chamber measures how well the metallic enclosure isolates the internal test volume from external interference. If shielding is inadequate, ambient noise directly contaminates the measured emissions of equipment under test, distorting or even invalidating EMC test results.
1. Standard Series Overview 📌
The BS EN 50147 series comprises two parts. BS EN 50147-1 specifies the general requirements and method for measuring shield attenuation; BS EN 50147-2 is the complementary part covering engineering application and acceptance details. The series was developed by CENELEC technical committees and published in the UK by BSI Technical Committee GEL/210 (Electromagnetic Compatibility), consolidating earlier scattered specifications for electromagnetic shielded rooms.
It is important to note that the anechoic chamber is the core facility for radiated emission (RE) and radiated immunity (RS) testing. The shielding effectiveness of the metallic enclosure, the performance of the absorber lining, and the quality of ancillary parts such as power-line filters and ventilation waveguides jointly determine whether the chamber meets the required ambient noise and site-attenuation levels. BS EN 50147 focuses on how to measure and verify these properties.
2. Series Composition and Technical Points 📋
| Standard | Title | Main Content |
| BS EN 50147-1:1997 | Anechoic chambers — Part 1: Shield attenuation measurement | Defines measurement method, test equipment, frequency range, procedure, and evaluation of shield attenuation (readable text) |
| BS EN 50147-2:1997 | Anechoic chambers — Part 2: Supplementary requirements (scan version) | Engineering application, acceptance, and implementation details (scan version, written from public-domain knowledge) |
⚠️ BS EN 50147-2 in this series is a scan version; its body is written from public-domain knowledge for reference only. For formal acceptance and compliance decisions, refer to the official version published by BSI.
3. Key Technical Requirements 🔬
3.1 Definition and magnitude of shield attenuation
Shield attenuation describes the loss experienced by electromagnetic waves passing through a shield, normally expressed in decibels (dB). The higher the value, the better the isolation. The standard requires transmit and receive antennas to be placed inside and outside the chamber respectively, and the attenuation to be calculated from the difference in received signal with and without the shield. Since the result varies with frequency, measurements are taken point by point at specified frequencies to build a complete attenuation-versus-frequency curve.
3.2 Measurement frequency range
Shield attenuation measurement spans a broad frequency range, from tens of kilohertz to the gigahertz region. At low frequencies the magnetic-field component dominates, so measurement targets magnetic shielding; at high frequencies the electric-field and plane-wave components dominate and the emphasis shifts to electric-field and plane-wave shielding. Different antenna types and sizes are used in different bands, and the antenna-to-shield distance must be controlled to satisfy near-field or far-field conditions.
3.3 Test layout and procedure
- Transmit and receive antennas: placed inside and outside the chamber with matched polarisation to avoid polarisation-mismatch errors.
- Swept and point-frequency measurement: measure point by point at the specified frequency steps while recording source level, received level, and antenna factors.
- Data processing: shield attenuation is the net difference of transmit/receive power levels after corrections for antenna factor, distance, and instrumentation, usually averaged over multiple readings.
- Calibration: calibrate the signal source, receiver, and antenna system before measurement so the amplitude/phase characteristics of the chain are known.
4. Engineering Practice 🛠️
- 🔍 Chamber acceptance: a new or refurbished anechoic chamber must be acceptance-tested to BS EN 50147 to confirm the shielding performance at every band meets the design specification before it is put into service.
- 📊 Ambient-noise assessment: after commissioning, periodically check the ambient noise level to ensure it stays a sufficient margin below the product emission limits, preventing ambient noise from inflating measured emissions.
- 🔧 Shielding integrity: chamber doors, power filters, ventilation waveguides, and feed-through interfaces are the weakest points; keep them sealed and well grounded during normal use.
- 📋 Link to EMC standards: BS EN 50147 results verify that the chamber satisfies the background-noise and site-attenuation requirements of the relevant EMC product standards (e.g., radiated emission and radiated immunity tests).
5. Frequently Asked Questions ❓
❓ Q: What is the difference between an anechoic chamber and a plain shielded room?
A: A shielded room provides only metallic shielding; its walls usually have no absorber. An anechoic chamber adds absorber lining to the shielded enclosure to simulate free-space (or semi-free-space) propagation, making it suitable for radiated emission and radiated immunity tests. Both need adequate shielding, but the anechoic chamber additionally requires sufficient absorption performance.
❓ Q: Is shield-attenuation measurement the same as site-attenuation measurement?
A: No. Shield attenuation measures how well the enclosure isolates against external waves (dB), i.e., how “leak-free” the chamber is. Site attenuation measures the transfer loss between transmit and receive antennas along the propagation path inside the chamber and relates to the absorption of the absorber lining. EMC testing normally requires both to be satisfied.
❓ Q: Why is magnetic shielding measured mainly at low frequencies?
A: Low-frequency electromagnetic waves are dominated by the magnetic-field component, which has a long wavelength and high penetrating power, making it hardest to shield. Consequently, low-frequency tests focus on magnetic shield attenuation, while high-frequency tests focus on electric-field and plane-wave attenuation.
❓ Q: How does excessive chamber background noise affect product testing?
A: Excessive ambient noise superimposes on the emissions of the product under test, making the measured values too high and potentially causing compliant products to be judged non-compliant. It is normal practice to require the chamber ambient noise to be at least 6 dB below the product emission limit.